4 Troubleshooting
Doc. 065215-02 9/09 59
Liquid leaks at junction between pump head and pump housing
Replace the piston seal (see Section 5.7
).
4.4 Pump Does Not Start
No power (front door Power LED indicator fails to light)
Check that the power cord is plugged in.
Check the main power fuses and replace if needed (see Section 5.13
).
No communication between ICS-900 and Chromeleon or Chromeleon
Xpress (
Link LED on rear panel fails to light)
The USB cable is not connected correctly. For connection instructions, see the
ICS-900 Ion Chromatography System Installation Instructions (Document
No. 065214), provided on the Dionex Reference Library CD-ROM (P/N
053891) and in the ICS-900 Ship Kit (P/N 067768).
4.5 No Flow
Pump not primed
Prime the pump (see Section 3.6
).
Broken pump piston
Replace the piston (see Section 5.8
).
4.6 Erratic Flow/Pressure Reading
Pump needs priming
Prime the pump (see Section 3.6
).
Damaged piston seal
Replace the piston seal (see Section 5.7
).
Dirty pump check valve
Clean or replace the check valve (see Section 5.6
).