ICS-900 Operator’s Manual
40 Doc. 065215-02 9/09
3.11.2 Manually Processing Samples
To manually process a sample, select operating parameters and
commands from the Chromeleon or Chromeleon Xpress panel tabset.
Commands are executed as soon as they are entered.
Summary of Manual Sample Processing
1. Complete the instructions in Section 3.2 through Section 3.10 to
prepare the ICS-900 for operation and to prepare the sample for
processing.
2. Load the sample into the injection valve sample loop with an
autosampler (see Section 3.11.4
) or a syringe (see Section 3.11.5).
3. On the ICS-900 Control panel, click the Autozero
button.
4. On the Sequence Control panel, click the Acq On button (or click
the Acquisition On/Off button on the Chromeleon toolbar).
5. Switch the injection valve to the Inject position. The method used to
switch the injection valve depends on whether you are using an
autosampler (see Section 3.11.4
) or a syringe (see Section 3.11.5).
6. The signal plot is displayed on the ICS-900 Control panel (see
Figure 3-9
). Monitor the chromatogram and when sample data has
been collected, click the Acq Off button on the Sequence Control