ICS-900 Operator’s Manual
64 Doc. 065215-02 9/09
Microbore systems: Use one or two red backpressure coils (P/N 045878)
supplied in the microbore tubing kit (P/N 052324).
Install the tubing between the cell outlet and the regenerant bottle inlet. For
connection instructions, see the ICS-900 Ion Chromatography System
Installation Instructions (Document No. 065214), provided on the Dionex
Reference Library CD-ROM (P/N 053891) and in the ICS-900 Ship Kit
(P/N 067768).
Insufficient system equilibration following changes to operating
parameters; especially apparent when operating at high sensitivities
Allow a longer system equilibration time (up to 3 hours) before starting to
inject samples.
Incorrect suppressor operating conditions
Refer to the suppressor manual for troubleshooting information. Suppressor
manuals are included on the Dionex Reference Library CD-ROM
(P/N 053891).
Cell above or below temperature
Contact Dionex for assistance.
Damaged piston seal
Replace the piston seal (see Section 5.7
).
DCR backpressure tubing not installed or incorrectly installed
Install the tubing on the end of the suppressor waste line. For connection
instructions, see the ICS-900 Ion Chromatography System Installation
Instructions (Document No. 065214), provided on the Dionex Reference
Library CD-ROM (P/N 053891) and in the ICS-900 Ship Kit (P/N 067768).
Flow Rate Number of Coils
0.3 to 0.5 mL/min
1 (red)
Less than 0.3 mL/min
2 (red)