eMechanical movement check
Execute 20 times mechanical movement checks to make sure whether the washing block of
rinsing mechanism nozzle abrases the reaction cuvette or not and each mechanism runs normally
or not.
fRinse reaction cup+ ISE
Select rinsing reaction cuvette in “Maintenance” interface, and execute rinsing reaction cuvette +
ISE if ISE equipment is collocated.
gLight quantity check
Light quantity result should be attached to installation check report with its value no more than
18000.
hCup blank test
No.1 cup blank value should be within 18000, and 2-120 reaction cup check value should be
within 18000 ±800.
2.6 Clinical item test
Edit chemical parameters; register reagent info.; testing rate assay ALT, point assay, two-point
rate BUN; calculate the difference of parameter and the result of test should be attached to
installation check report.
2.7 Train Medical Personnel
2.8 Fill the Installation Check Report Detailedly.