The injection pump can be seen in figure.
Observe whether the injection pump is leaking.
If so, check the leakage causes, and check the pipeline and connector timely.
7.2.2 Check/rinse sample & reagent probe
In online status, click “Instrument resetting” in “Maintenance”, and instrument
When cleaning samle&reagent probe, carefully observe whether the outflow of
sample&reagent probe internal wall is continuous, whether the direction of flow
is consistent with the sample&reagent probe and the outflow of external wall is
continuous, and whether water volume is normal.
If not normal, clean sample&reagent probe(refer to 7.3.1)
If still not normal, check the corresponding liquid line channel; check whether
water supply of water tank and water pressure is normal.
7.2.3 Rinsing stirring rod
In online status, click “Instrument resetting” in “Maintenance”, and instrument
When cleaning, carefully observe whether the stirring rod works normally, If not
normal, check the corresponding liquid line channel, check whether water supply
of water tank and water pressure is normal.
7.2.4 Rinse rinsing mechanism
In online status, click “Rinse reaction cuvette” in “Maintenance”, and instrument
executes reaction cuvette rinsing.