68
Caution
Sample&reagent probe is precisely processed to ensure the sample and
reagent adding precision. If the probe point is damaged or bent, checking
sample reagent probe is a must, or no guarantee can be made for test
precision.
7.7.4 Rinse sample&reagent probe
When it is found that the water level of rinsing bath is too high when rinsing sample&reagent probe
because of no discharging available, which may be caused by the blocked leaking hole. Cleaning sample
&reagent rinsing bath is necessary.
Warning
Please be careful to avoid being scratched by probe.
Biological contamination danger:
In operation, please put on gloves, work cloths, and put on protective glasses
for the best
1
Make sure the analysis part power is switched off.
2
Lift the rotating arm of sample&reagent probe by hands to its top position,
and rotate its rotating arm to keep sample&reagent probe away from rinsing
bath for convenient operation.
3
Infuse about 1ml alkaline detergent of 0.5% (V / V) sodium hypochlorite or 84
disinfectant into rinsing bath for 10 minutes.
4
Switch on the power supply of analysis part
5
Lift the rotating arm of sample&reagent probe by hands to its top position,
and rotate its rotating arm to keep sample&reagent probe above the rinsing
bath of sample&reagent probe.
Caution
Please rotate the sample&reagent probe to the top of sample&
reagent probe rinsing bath after clean rinsing bath of sample
&reagent probe.
6
Select and execute “Instrument resetting” after enter “Maintenance-routine
maintenance”, and the system will reset sample &reagent probe and rinsing
bath will be rinsed with deionized water automatically. Observe the outflow of
reagent&sample probe rinsing bath.
7.7.5 Rinse stirring rod