Chapter 2 Installation
2.1 Space Requirement
To make sure the space of maintenance, operation and repair, please follow the instruction as
below:
Space between left (right) side of analyzer and the wall should 50cm
Space between rear board of analyzer and the wall should 50cm
Space in front of analyzer should100cm
Make sure there is enough space for waste device and purified water equipment.
2.2 Power supply requirement
Power supply: AC220V±22V 50Hz/60Hz
Power:
650VA
Circuit breaker: 250V, 20A
A well grounded power supply socket is a must. Large electrical appliance such as air condition,
refrigerator, even cannot use the same socket with analyzer.
! Warning:
Incorrect grounding may cause electric shock or instrument damage.
Input voltage should conform to requirement. 3KVA-line UPS power supply is advised.
2.3 Environment requirement
Working temperature: 15℃~32
Relative humidity: 32 %~85
Atmospheric pressure: 76kPa106kPa
Environment should with no dust, vibration, major noise source and power interference
Do not put the analyzer in the vicinity of brush motor, flicker fluorescent tube and other
constant on-off electrical equipment.
Hard and flat ground is a must for the instrument.
Avoid direct sunlight, do not put the analyzer in front of heat source and wind source
Keep good ventilation.
! warning
Normal running and accurate result can not be guaranteed if instrument works beyond the
requirements mentioned above.