63
Biological contamination danger:
In operation, please put on g
loves, work cloths, and put on
protective glasses for the best.
Do not dispose the gauze used to clean sample&reagent
probe.rinsing bath at your own will, please follow the relevant
provisions for proper disposal.
1
Make sure the analysis part power is switched off.
2
Lift the rotating arm of sample&reagent probe by hands to the its top position, and
rotate its rotating arm to keep reagent&sample probe away from rinsing bath for
convenient operation.
3
Clean the inside and appearance of sample&reagent probe rinsing bath with clean
cotton stick.
4
After cleaning, lift the rotating arm of sample&reagent probe to the top position,
and rotate the rotating arm of sample&reagent probe to locate the probe to the top
of the rinsing bath.
Caution
After the work of sample&
reagent probe surface rinsing,
please make sure to rotate reagent&sample probe to the top
of reagent&sample probe rinsing bath.
5
Switch on the power of analysis part and wait 30 seconds, enter the "maintenance -
routine maintenance" column to implement “instrument resetting", the system will
automatically reset the sample&reagent probe.
7.4.2 Rinse stirring rod rinsing bath
Warning:
Please be careful to avoid being scratched.
Biological contamination danger:
In operation,
please put on gloves, work cloths, and put on
protective glasses for the best.
Do not dispose the gauze used to clean stirring rod at your own
will, please follow the relevant provisions for proper disposal.
1
Make sure the analysis part power is switched off.
2
Lift the stirring arm to the top position by hand to one side of rinsing bath.
3
Wipe stirring rod with clean soft gauze.