EasyManua.ls Logo

Dirui CS-T240 - AD Collector

Dirui CS-T240
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
121
144-2
refrigeration
system
abnormal
refrigeration
current
abnormal
1observe whether current value displayed on
digital pipe of refrigeration system is normal.
2Make sure which current is abnormal, and cope
with it after check the corresponding abnormal Peltier
and circuit board.
144-3
refrigeration
system
abnormal
refrigeration
chip
abnormal
1observe which chip is abnormal
2cope with the corresponding abnormal chip
144-5
refrigeration
communicatio
n abnormal
refrigeration
status
abnormal
1check the communications wiring of refrigeration
board and main board
2check the communications interface circuit of
refrigeration board
3check the communications interface circuit of
main board
145-1
The 1
st
line
refrigeration
chip
malfunction
The 1
st
line
refrigeration
current <5A
Check the 1st line refrigeration Peltier and cooling
chip
145-2
The 2nd line
refrigeration
chip
malfunction
The 2nd
line
refrigeration
current <5A
Check the 2nd line refrigeration Peltier and
refrigeration chip
9.18 AD Collector
Alarm
Code
Description
Detailed
description
Solution
146-1
The 1
st
line
AD collector
The
measuring
value of the
Check the AD collection board and preamp board

Table of Contents