Switch on the power of analysis part and wait 30 seconds, enter the "maintenance
- routine maintenance" column to implement “instrument resetting", the system
will automatically reset the sample&reagent probe and rinse them with deionized
water.
7.3.3 Sample/ barcode window rinsing
Caution:
Do not gaze scanning laser light, or it may cause eyes injury
Make sure the analysis part power is switched off.
Remove the reagent sample disk cover, and then remove the sample&reagent disk.
Wipe the scanning glass window lightly with gauze dipped with deionized water.
Remount the sample&reagent disk and cover them.
Switch on the analysis part and wait 30 seconds, the system will reset
automatically.
7.3.4 Rinse reaction cuvette
The contamination of sample&reagent probe, stirring rod and reaction cuvette will affect the
accuracy of measurement. The reaction cuvette requires intensive rinsing.
Place 70 ml detergent (1mol of hydrochloric acid or 0.5% NaOH solution, use
the two types of solution alternatively, one week one solution) at the 45th
detergent position of sample&reagent disk.
Click “Maintenance” functional key to enter “System maintenance” menu,
and select “Rinse reaction cuvette” to execute.
7.4 Monthly maintenance
7.4.1 Rinse sample&reagent probe rinsing bath
Warning:
Please be careful to avoid being scratched by probe.