Note:
Sample&reagent probe is precisely processed to ensure the sample and
reagent adding precision. If the probe point is damaged or bent, checking
sample&
reagent probe is a must, or no guarantee can be made for test
precision.
7.7.2 Clean sample&reagent probe
Put a stainless steel wire through the sample&reagent probe point to clean the
impurity in the probe.
Caution:
Sample&reagent probe is precisely processed to ensure the sample and
reagent adding precision. If the probe point is damaged or bent, checking
reagent&
sample probe is a must, or no guarantee can be made for test
precision.
7.7.3 Install sample&reagent probe
Warning
Please be careful to avoid being scratched by probe.
Biological contamination danger:
In operation, please put on gloves, work cloths, and put on protective
glasses for the best
Dismounting sequence is opposite to that of sample&reagent probe.
Warning
Please be careful to avoid being scratched by probe.
Biological contamination danger:
In operation, please put on gloves, work cloths, and put on protective glasses
for the best.
Do not dispose the gauze used to clean sample&reagent probe at your own
will, please follow the relevant provisions for proper disposal.