66
1
Make sure the analysis part power is switched off.
2
Remove sample&reagent disk cover and then sample&reagent disk
3
Lift the rotating arm of sample&reagent probe by hands to its top position, and
rotate its rotating arm to keep sample&reagent probe away from rinsing bath for
convenient operation.
4
Hold shell claw of probe rotating arm with fingers and lift to remove.
5
6
Loosen pipeline interface
7
warning
Carefully place dismounted sample reagent&probe and prevent it
scratching human body and sample&reagent probe damage.
Note
Take out sample reagent&
probe from the rotating arm and be
careful to operate to avoid the damage of probe point caused by
touching rotating arm.