7.5 Every 6 months maintenance
7.5.1 Check light source lamp
Lamp light source of optical system will gradually be aged in use, and will cause an increase in noise
during measurement. If the cuvette blank and light source intensity attenuation is out of range or the
working time of light source lamp accumulates over 2000 hours, the light source lamp should be
checked.
Caution:
Please use consumables recommended by DIRUI company, using other
consumables may cause system performance degradation.
Do not touch the light source lamp shell surface and lens in front of
the light source lamp by hand, because it may change the characteristics
of the light source. If you accidentally make noodle stained with filth,
absorbent cotton dipped by absolute alcohol can be used to clean it.
of reaction cuvette. Loosen the setscrews of reaction disk and remove the
reaction disk. Place the reaction cuvette at dry and clean position. Loosen the
two fixing wire connecting poles of halogen and remove down-lead.