65
3
Loosen the two screws fixing light source seat to remove halogen lamp.
4
Mount a new halogen lamp according to the above opposite steps; pay attention to
tighten the screws. The cooling rubber hose in the lamp room can not be twisted and
down-lead can not be loosed or cocked.
5
Remount the reaction disk, the reaction cuvette and rinsing mechanism; switch on
the power supply of analysis part. After standby mode, single-click “Next” in
“System maintenance” window; infuse purified water into reaction groove. After
instrument standby mode, execute light quantity check function. Check the back of
halogen lamp if the light quantity conforms to the requirement to start test.
7.6 Every year maintenance
7.6.1 Check water of cooling system
7.7 Irregular check
7.7.1 Rinse sample reagent probe
If the water flow is not normal when rinsing sample&reagent probe, sample&reagent probe,the probe
may have been blocked and cleaning is needed to reagent&sample probe.
Warning:
Please be careful to avoid being scratched.
Biological contamination danger:
In operation, please put on gloves, work cloths, and put on
protective glasses for the best.