Switch on the power of analysis part and wait 30 seconds, enter the
"maintenance - routine maintenance" column to impl
resetting", the system will automatically reset the sample&reagent probe and
rinse them with deionized water.
7.3.2 Rinse stirring rod
Make sure the analysis part power supply is switched off.
Lift the stirring rod by hands to the top position, and rotate its rotating arm to a
position for convenient operation.
When cleaning, do not touch the stirring rod surface directly to
prevent scratch; avoid too much hand force to prevent deformation
of the sample&reagent probe.
Acidic and alkaline detergent can be used alternatively, for instance,
acidic detergent is used at previous time maintenance, use alkaline
detergent at this time maintenance.
Wipe the surface of stirring rod lightly with cotton stick moisturized with
alcohol, especially the point of probe, until no impurities left at all.
Wipe stirring rod with the gauze dipped with deionized water
After cleaning, lift the rotating arm of stirring rod to the top position, and rotate
the rotating arm of stirring rod to locate the stirring rod to the top of the rinsing
bath.
Biological contamination danger:
In operation, please put on gloves, work cloths, and put on protective
glasses for the best.
Do not dispose the gauze used to clean stirring rod at your own will,
please follow the relevant provisions for proper disposal.