60
1
Make sure the analysis part power supply is switched off.
2
Lift the rotating arms of samle&reagent probe by hands to the top position, and
rotate it to the top of sample&reagent storehouse for convenient operation.
3
Caution:
When cleaning, do not touch directly the probe surface to
prevent probe scratch; avoid too much hand force to prevent
deformation of the samle&reagent probe.
Note:
Acidic and alkaline detergent can be used alternatively, for
instance, acidic detergent
is used at previous time
maintenance, use alkaline detergent at this time maintenance.
Wipe the external walls of sample&reagent probe lightly with cotton
stick moisturized with alcohol, especially the point of probe, until no
impurities left at all.
4
Wipe sample&reagent probe with the gauze dipped with deionized water.
5
After cleaning, lift the rotating arms of sample&reagent probe to the top
position, and rotate the rotating arm of sample&reagent probe to locate the
sample&reagent probe above the rinsing bath of sample&reagent probe.
Caution:
After cleaning the surface of a sample&reagent probe, please
make sure sample&reagent probe must be rotated to the top
of sample&reagent probe rinsing bath.