101
9.14 Incubation bath malfunction analysis
Alarm
Code
Description
Detailed
description
Solution
20-1
Incubation bath
temperature
abnormal
The
temperature of
incubation
bath is above
45
(1) Make sure the cooling fan of the instrument is rotating
normally
(2) Check the temperature sensor is working normally
(3) Check the connector part of the temperature sensor on
the main control board
(4) Replace the main control board or the temperature
sensor
20-2
Incubation bath
temperature
abnormal
The
temperature of
the incubation
bath is out of
the scope of
37℃ ±5℃
(Only check
when the
instrument
under
operation)
(1) Make sure the room temperature is in the scope of
15-32
(2) Make sure the cooling fan of the instrument is rotating
narmally
(3) Make sure the water is cycling in the incubation bath
(4) Replace main control board or the temperature sensor
9.15 ISE Failure Analysis
0-31 Note
ISE is over
measuring lowe
limit
ISE QC is
over
measuring
lower limit
the current
QC has no
result.
1Please check whether the Control
volume, reagent volume is adequate, and
whether the position is correct
0-32 Note
ISE is over
measuring
upper limit
ISE QC is
over
measuring
upper limit
the current
QC has no
1Please check whether the Control
volume, reagent volume is adequate, and
whether the position is correct