102
result.
0-33 Note
ISE is over
measuring
lower limit
ISE
Calibration
is over
measuring
lower limit
the current
Calibration
has no
result.
1Please check whether the Calibrator
volume, reagent volume is adequate, and
whether the position is correct
0-34 Note
ISE is over
measuring
upper limit
ISE
Calibration
is over
measuring
upper limit
the current
Calibration
has no
result.
1Please check whether the Calibrator
volume, reagent volume is adequate, and
whether the position is correct
0-35 Note
ISE is over
measuring
range
ISE sample
is over
measuring
upper or
lower limit
no test
result.
1Please check whether the sample
volume, reagent volume is adequate, and
whether the position is correct
19-1 Note
Electrolyte
function stops
Electrolyte
system
stops
working
due to the
issued
alarm.
Note: the
stop status
of adding
sample
means to
restart.
1Enter into system maintenance, and
execute "Resetting”, and then execute
“Mechanical movement check”
37-1 Note
ISE reference
liquid is
inadequate
Remaining
volume of
reference
solution is
inadequate
(less then
the user
designed
1add ISE reference solution
2Enter into system maintenance
window, and execute "ISE reference
solution reagent pipeline rinsing”
3Execute ISE calibration