volume of
internal
standard
solution is
inadequate
(less then
the user
designed
volume)
(1)add ISE internal standard solution
(2)Enter into system maintenance, and
execute "ISE internal standard solution
reagent pipeline rinsing”
(3)Execute ISE calibration
39-1 Note
ISE diluent is
inadequate
volume of
diluent is
inadequate
(less then
the user
designed
volume)
(1)add ISE diluent
(2)Enter into system maintenance, and
execute "ISE diluent reagent pipeline
rinsing”
(3)Execute ISE calibration
60-1 Note
ISE LEVEL
error
In the
5-measurin
g point
potential of
internal
standard
solution, the
average
value
(EAV)of
three
potentials is
over the
following
range
(internal
standard
solution)K:
-90.0mv≤E
AV≤-10mv
(1) Enter into system maintenance
window, and execute "ISE check."
(2) please refer to " electrolyte device
maintenance " in the "user manual" for
detail.
60-2 Note
ISE LEVEL
error