104
potential of
internal
standard
solution, the
average
value
(EAV)of
three
potentials is
over the
following
range
internal
standard
solutionK
-90.0mv≤E
AV-10mv
(2) please refer to " electrolyte device
maintenance " in the "user manual" for
detail.
60-3 Note
ISE LEVEL
error
In the
5-measurin
g point
potential of
internal
standard
solution, the
average
value
(EAV)of
three
potentials is
over the
following
range
internal
standard
solution
Cl
80.0mv≤EA
V≤160mv
(1) Enter into system maintenance, and
execute "ISE check."
(2) please refer to " electrolyte device
maintenance " in the "user manual" for
detail.
61-1 Note
ISE Noise
error
In the
5-measurin
g point
potential of
internal
standard
(1) Enter into system maintenance, and
execute "ISE check."
(2) please refer to " electrolyte device
maintenance " in the "user manual" for