solution, the
difference
FI of
the
maximum
and
minimum is
within
following
range
Na
0.7mv≤|FIV
2-FIV
4|
detail.
61-2 Note
ISE Noise
error
In the
5-measurin
g point
potential of
internal
standard
solution, the
difference
FIV of
the
maximum
and
minimum is
within
following
range
(internal
standard,
sample)
K
1.0mv≤|FIV
2-FIV
4
(1) Enter into system maintenance, and
execute "ISE check."
(2) please refer to " electrolyte device
maintenance " in the "user manual" for
detail.
61-3 Note
ISE Noise
error
In the
5-measurin
g point
potential of
internal
standard
solution, the
difference
FIV of
the
(1) Enter into system maintenance, and
execute "ISE check."
(2) please refer to " electrolyte device
maintenance " in the "user manual" for
detail.