59
2
When rinsing, carefully observe rinsing probe working and whether probe infusing
is normal and assimilating is completely.
If infusing abnormal, check pressure value of water infusing pressure gauge
If assimilating abnormal, check pressure value of assimilating vacuum
7.2.5 Check waste connection and discharging
7.2.6 Rinse instrument surface
1
Check whether liquid waste disposal system is normal every day, and
maintain waste liquid pipe is not bent and discharges smoothly and high
and low concentration waste liquid are disposed properly (refer to local
standards of dispose waste liquid).
7.2.7 Check printer and printing paper
7.3 Weekly maintenance
7.3.1 Rinse sample reagent probe
Warning:
Please be careful to avoid hands from being scratched
Biological contamination danger:
In operation, please put on gloves, work cloths, and put on protective glasses
for the best.
Do not dispose the gauze used to clean sample probe at your own will, please
follow the relevant provisions for proper disposal.
1
Biological contamination danger:
During waste water operation, please put on gloves, put on work
clothes and if necessary, wear protective glasses.
Check whether liquid waste disposal system is normal every day, and maintain
waste liquid p
ipe is not bent and discharges smoothly and high and low
concentration waste liquid are disposed properly (refer to local standards of
dispose waste liquid).
1
Check printer power supply indicator, preparation indicator and printing paper
daily.