EasyManua.ls Logo

Dirui CS-T240 - Page 120

Dirui CS-T240
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
120
143
-45
Continuous
emergence of
dirty cups
Continuous
emergence of 5
dirty cups
Check whether light source, water quality of
incubation bath, reaction cup and counting light
sensor of reaction disk are normal; test data
collecting board lines by using the testing program
of data collecting board.
143
-46
AD data
malfunction
AD data
mixed
Execute the main board debuggin
g program to
monitor whether the communications of main
control board and AD board is normal, and analyze
error; check whether counting light sensor is
normal.
143
-47
IES
malfunction in
testing
ISE measuring
internal
standard liquid
failure
Use lower machine monitoring software to monitor
error information, and refer to the related error
information of ISE board to debug.
143
-48
Version
number
reading
overtime
Version number
reading
overtime
Execute the main board debugging program to
monitor whether the communications is normal, and
analyze error; check whether relevant electrical
units of sub-
boards communications and control
board are normal.
9.17 Refrigeration system
Alar
m
Code
Description
Detailed
description
Solution
144-1
refrigeration
system
abnormal
refrigeration
time
abnormal
(1)check whether sample&reagent disk cover is
covered, whether the ambient temperature is in line
with environmental requirements of instrument.
(2) observe the temperature displayed on digital pipe
of refrigeration system and Peltier current value are
normal.
If abnormal: Please check Peltier and refrigeration
circuit board

Table of Contents