143
-36
ISE
malfunction in
testing
ISE malfunction
when testing,
follow-up ISE
stops
Observe whether there is ISE alarm information;
Use lower machine monitoring software to monitor
error information, and refer to the related error
information of ISE board to debug.
143
-37
Gear pump
failure
Gear pump
pressure low
Check gear pump and pressure sensor of sample
pipeline are normal. Check whether the read value
of pressure sensor which is running is more than
2500 by using main control board debugging
software; check whether relevant electrical units of
pressure sensor and control board are normal.
143
-42
Sending
reagent
mapping
information
failure
Sending reagent
mapping
information
failure, adding
reagent may
fail.
Refer to instrument module error analysis
143
-43
cooling water
overtime
When the water
tank
temperature is
over 36.5
degrees, add
cold water into
the tank into to
cool down.
Temperature
could not drop
to 35.5 degrees
in one minute,
which means
cold water
temperature is
too high.
Execute the main board debugging program to
observe whether the display of temperature is
normal and room temperature is high; check the
water supply pipeline of water tank is normal; check
whether relevant electrical units of temperature
sensor and control board are normal.
143
-44
Analyzer
module
malfunction
Malfunction
among
modules
Execute the main board debugging program to
monitor whether the communications of main board
and sub-boards are normal; check whether relevant
electrical units of sub-boards communications and
control board are normal.