118
information of reaction disk to debug.
143
-28
Waste liquid
bottle full
Waste liquid
bottle full
Check whether waste liquid bottle is full; check
whether the sensor in waste liquid bottle is normal;
check whether relevant electrical units of the floater
in waste liquid bottle and control board are normal.
143
-29
Floater switch
error
Switch
malfunction,
high-level
floater detects
the signal, but
low-level floater
fails.
Check whether the floaters of high and low liquid
level and the signal cable connecting floater to main
control board are normal; check whether relevant
electrical units of floater detection and control board
are normal.
143
-30
Reagent level
scanning
overtime error
Reagent level
scanning
overtime error
Use lower machine monitoring software to monitor
error information, and refer to the related error
information of reagent board to debug.
143
-31
Vacuum Pump
Failure
Vacuum Pump
negative
pressure low
Check whether the vacuum pump and vacuum
pump switch are normal; check whether relevant
electrical units of vacuum pump switch and control
board are normal.
143
-32
Sample
barcode
scanning
overtime in
testing
Sample barcode
scanning
overtime in
testing
Use lower machine monitoring software to monitor
error information, and refer to the related error
information of sample board to debug.
143
-33
ISE reset
failure
ISE detects
malfunction in
resetting;
resetting
failed.
Observe whether there is ISE alarm information;
Use lower machine monitoring software to monitor
error information, and refer to the related error
information of ISE board to debug.
143
-34
ISE check
maintenance
movement
overtime
ISE check
maintenance
movement
overtime
Observe whether there is ISE alarm information;
Use lower machine monitoring software to monitor
error information, and refer to the related error
information of ISE board to debug.
143
-35
ISE pipeline
rinsing
overtime
ISE pipeline
rinsing is over
specified time
Observe whether there is ISE alarm information;
Use lower machine monitoring software to monitor
error information, and refer to the related error
information of ISE board to debug.