Reaction bath
liquid level
malfunction
Reaction bath
liquid level
malfunction
Check whether the liquid level detector of reaction
bath is clean, whether there is water in reaction bath,
whether water level reaches the position where the
detector can detect, and check liquid level detector
in reaction bath is working normally; check whether
relevant electrical units of liquid level detection and
control board are normal.
143
-14
Bar code
scanning
overtime error
Bar code
scanning
overtime error
Use lower machine monitoring software to monitor
error information, and refer to the related error
information of reagent&sample disks to debug.
143
-15
Pipeline
exhaust
overtime error
Pipeline exhaust
overtime error
Execute the main board debugging program to
monitor, and re-execute pipeline air exhausting and
observe whether air exhausting is carried out by
reagent and sample injection pumps and the
implementation is complete. Use lower machine
monitoring software to monitor error information,
and refer to the related error information of reagent
&sample disks to debug.
143
-16
sample
&reaction disk
start failure
sample
&reaction disk
start failure
Use lower machine monitoring software to monitor
error information, and refer to the related error
information of reaction disk to debug.
143
-17
Reaction disk
stop failure
Reaction disk
stop failure
Use lower machine monitoring software to monitor
error information, and refer to the related error
information of reaction disk to debug.
143
-18
Sample&
reagent probe
blocked
Sample&
reagent probe
blocked
Maintenance operation of the sample&reagent
probe. Check whether the pressure sensor and
magnetic valve of sample&reagent probe pipeline
are normal; check whether relevant electrical units
of pressure detection and control board are normal.
143
-19
Previous
sample adding
failure
Previous sample
adding failure
Use lower machine monitoring software to monitor
error information, and refer to the related error
information of sample disk to debug.
143
-20
Previous
reagent adding
failure
Previous
reagent adding
failure
Use lower machine monitoring software to monitor
error information, and refer to the related error
information of reagent to debug.
143
-24
Previous
stirringfailure
Previous stirring
failure
Use lower machine monitoring software to monitor
error information, and refer to the related error

Table of Contents