116
failure.
143
-4
Reaction disk
reset failure
Reaction disk
detects
malfunction
when resetting,
Reaction disk
reset failure
Lower machine monitoring software monitors error
information, and analyze the reaction disk
board-related failure to eliminate the error
143
-5
Sample&
reagent disk
reset failure
Sample&
reagent disk
detects
malfunction
when resetting,
sample&
reagent disk
fails to reset
Use lower machine monitoring software to monitor
error information, and refer to the related
information of sample&reagent disk board to debug.
143
-8
Water
discharging
failure of
reaction bath
Water
discharging
failure of
reaction bath
Check whether the indication of reaction bath liquid
level detector is right; check whether there is
remaining water in reaction bath, and check reaction
bath water outlet pipeline if there is remaining water
in it; check whether relevant electrical units of
liquid level detection and control board are normal.
143
-9
Adding
detergent
overtime error
Sample&
reagent probe
fails to add
detergent
completely in
the specified
time
Connect the main control board debugging program
to electrify the machine; observe machine
electrifying flow; observe whether the 6 times the
normal cleansing movement of detergents reagent is
finished in the phase of adding detergents.
Corresponding control board of reagent&sample
probe should be repaired if the reagent&sample
probe did not achieve normal movement.
143
-10
liquid level
detection
failure
fails to detect
liquid level
when adding
detergent
Refer to liquid detection error analysis to debug
143
-12
Liquid level
detection
failure of
reaction bath
Liquid level
detection failure
of reaction bath
Check the liquid level of reaction bath, and check
whether the liquid level detector of reaction bath
works normally; check whether relevant electrical
units of liquid level detection and control board are
normal.