69
If the stirring rod is damaged, please check stirring rod in accordance with following steps strictly.
Warning
Please be careful to avoid being scratched by probe.
Any touch is forbidden except at the knurling of it only by hand when
checking, and prevent any scratch on the flat part of stirring part.
Biological contamination danger:
In operation, please put on gloves, work cloths, and put on protective glasses
for the best
Please deal with removed stirring rod properly.
Caution
Please use consumables recommended by DIRUI company, using other
consumables may cause system performance degradation.
1
Make sure the analysis part power is switched off.
2
Prepare a new stirring rod and wipe the flat part of it with gauze or cotton stick
dipped with cleaning liquid, and then wipe it with gauze dipped with deionized
water.
3
Lift the rotating arm of stirring rod by hands to its top position, and rotate its
rotating arm for convenient operation.
4
Caution
When pulling out stirring rod, make sure the direction of force is
vertical to that of axis of rotating arm. Lateral force may damage
the axis or stirring rod.
Remove stirring rod after loosen the two fixing screws.
5
Prepare a new stirring rod, and wipe the front of the stirring rod with gauze
dipped with 2% -antibacterial detergent.