55
The instrument add calibrator into the reaction cuvette.The SIP injection pump aspirate the calibrator,
and pump it into the NaKCl electrode pipeline of the reaction tank. Test the electric potential on the
basis of reference electrode. The sample probe aspirates sample and diluent, and pump it into the
reaction cuvette. Test the electric potential, and the concentration can be calculated.
J2optocoupler port of aspirate nozzle elevator mechanism
J5motor port of proportioning pump
P13optocoupler port of proportioning pump
J61motor port of aspirate nozzle elevator mechanism
J62solenoid valve port
J64preamplifier board port
ISE preamplifier boardamplify electrode signal