EasyManua.ls Logo

Dirui CS-T240 - Page 71

Dirui CS-T240
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
71
2
Put on protective gloves to remove fixing screws.
3
Rotate reaction disk by hand and remove the reaction cuvette sequently. Take
out reaction cuvette while rotating it.
4
Rinse the new reaction cuvette dipped in with water; rinse inside and outside
of reaction cuvette and no scratch is allowed.
5
Rotate reaction disk by hands, and mount new reaction cuvette on reaction
disk and check the six sets reaction cuvette simultaneously.
6
Mount reaction disk with the opposite steps and make sure the fixing screw of
reaction disk is tight.
7
Switch on the power of analysis part