2、No change on multimeter reading
(1) Check conductors
Check whether the P02 plug 4,5,6 core well-connected, the
connection is connected at both ends of good
(2) Check the coupler signal
Check voltage between P02 plug 4,6 feet, if the voltage is
not 5v, check the connection plug is connected good. If the
voltage is 5v, check the J02 socket 5 feet of potential,
rising to peak when the potential was high or low potential.
If the potential is not normal, check the coupler
Malfunction 3:Stirring mechanism cannot reach the zero
position.
solution:Mechanical repair
1-3
Stirring
mechanism
abnormal
stirring
mechanism
can leave the
top when it
descents
In the CS-T240 software, enter into the "system
maintenance" interface after on-line, implement
"mechanical movement check", and observe the
stirring rod running status
Malfunction 1:stirring mechanism stops
solution:
1, Check stirring mechanism up-down movements,
mechanical repair is required if resistance is big.
2, check whether both ends of connector of the motor are
connected well.
3, check whether the conductivity of motor lead wire is
good.
4, check the motor drive module of circuit board is
working normally
Malfunction 2:stirring mechanism can reach the top but not
leave
Solution :the same to 1-1 Malfunction 2
1-4
Stirring
mechanism
abnormal
stirring
mechanism
fails to reach
the rinsing
bath when it
moves to
rinsing bath.
In the CS-T240 software, enter into the "system
maintenance" interface after on-line, implement
"mechanical movement check", and observe the
stirring rod running status
Malfunction 1:
stirring mechanism doesn’t move
Solution
1, Check stirring mechanism up-down movements,