Chapter 2: Installation
Operation Manual: Evolution Series
2
-
16
Temperature Control
1. Install paddle blades.
2. Raise drive assembly.
3. Fill each vessel with 900ml of DI water at 18-22°C.
4. Turn height adjust knob to “Paddles” position.
5. Lower drive assembly and paddles fully.
6. Cover all vessels with vessel covers provided.
7. Place a calibrated reference temperature probe, such as Distek TempChek, in
one of front center vessels.
8. Record vessel temperature and time.
9. Set rotation to 150RPM (see Controlling All Vessels on page 3-32).
10. Set operating temperature.
11. Highlight START HT/STIR button and press ENTER.
12. Monitor temperature and time until vessel temperature stabilizes.
13. Adjust temperature set-point so that vessel temperature stabilizes at required
vessel temperature (such as 37.0°C).
14. Record set-point required to achieve required vessel temperature.
15. Verify & record temperature calibration of calibration probe and each vessel
probe using external reference thermometer at approximately 37°C.
Time and Date
Verify that the system time and date are correct.
1. From the Main screen select SYSTEM SETUP.
2. Select SYSTEM CONFIG.
3. Verify that the time and date parameters are correct.
4. Make changes if necessary.