Chapter 4: Maintenance
Operation Manual: Evolution Series
General Information
1. Unit should be cleaned periodically by wiping with a slightly damp cloth. Do
not use any harsh chemicals or detergents.
2. Electronics require no customer attention. In the event of a malfunction,
please call the factory before attempting any repairs.
3. Turn off unit and unplug power cord. Remove drive cover. Apply small
amount of Dow Corning High Vacuum Grease to top and bottom of both drive
belts. (This should be done every 6 months.) Replace cover. Drive motor
and bearings are permanently lubricated. Belt tension is factory adjusted.
4. Clean guide posts every three months with alcohol and wipe dry with clean
cloth or paper towel. Do not lubricate posts.
5. Temperature calibration probe should be cleaned off and stored in holder on
left rear side of drive unit when not in use.
6. Vessel support plate requires extra care. Do not allow acids or buffers to
remain on the plate for extended periods of time. WIPE THE PLATE CLEAN
AFTER USE. DO NOT SCRATCH OR SCORE PAINTED SURFACE.
7. TCS Fuse Replacement:
110V Fast Blow 5x20mm – 8A 250V
230V Fast Blow 5x20mm – 5A 250V
Do not clean plastic and rubber parts with organic
solvents.