Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 133
Tests
d. Connect the hose (6) to the coaxial test adapter (3).
2. Touch the Tests... button.
3. Touch the System test button.
The checklist opens and the breathing system pressure is displayed in the lower
left-hand corner.
4. Press the O
2+ key and keep it pressed until the displayed breathing system
pressure no longer rises.
5. Release the O
2+ key.
6. Wait until the pressure stabilizes between 15 and 30 hPa. If the pressure drops
further, the hose is leaking.
Checking the outer hose:
1. Connect the breathing circuit as follows:
a. Connect the hose (1) to the inspiratory port (2).
b. Remove the coaxial test adapter (3) from the expiratory port (4).
c. Connect the elbow (5) to the hose (6).
d. Plug the hose (6) with the elbow connected on to the circuit plug (7).
e. Connect the hose (8) to the expiratory port (4).
2. Touch the Tests... button.
3. Touch the System test button.
The checklist opens and the breathing system pressure is displayed in the lower
left-hand corner.
4. Press the O
2+ key and keep it pressed until the displayed breathing system
pressure no longer rises.
5. Release the O
2+ key.
6. Wait until the pressure stabilizes between 15 and 30 hPa. If the pressure drops
further, the hose is leaking.
7. Exit the system test again. To do this, touch the Cancel button.
39555
3
5
8
6
4
1
2
7

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals