134 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Tests
For leaking hoses, the complete leakage test can be used to determine how large
the leakage value is and whether the hose can continue to be used.
Complete leakage test
Testing the inner hose:
1. Connect the breathing circuit as follows:
a. Connect the hose (1) to the inspiratory port (2).
b. Connect the coaxial test adapter (3) to the expiratory port (4).
c. Remove the elbow (5) from the hose (6).
d. Connect the hose (6) to the coaxial test adapter (3).
2. Touch the Tests... button.
3. Touch the Leakage test button.
The value determined for leakage is displayed as Leakage (mech. vent.). Note
down the value as the leakage of the inspiratory hose.
39554
3
5
4
6
1
2