Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 155
Alarms
10.3.2 Silencing the alarm tone
The alarm tone can be silenced for a maximum of 2 minutes.
Press the alarm silence key (1) below the screen, see page 17.
The symbol and the remaining time for the silenced alarm tone are displayed in
the header bar.
If the cause of the alarm persists, the alarm tone starts again immediately after the
alarm silence ends.
During the alarm silence, only new alarms with a higher alarm priority or a higher
internal priority number compared with the silenced alarm are acoustically signaled.
For further information see: "Alarm – Cause – Remedy", page 201.
10.3.2.1 Reactivating the alarm tone
Press the alarm silence key again.
10.3.3 Downgrading and acknowledging alarm messages
Some alarms can be downgraded to low priority (note) or they can be cleared
completely. The relevant alarms can be recognized in the table "Alarm – Cause –
Remedy" on page 201 by the following remedial messages:
36004
1
Remedial message Effect
Use "ALARM RESET" to downgrade
alarm priority.
Alarm priority is changed to low (note).
Use "ALARM RESET" to acknowl-
edge alarm.
Alarm is cleared.

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals