Alarms and Messages
VF4 Infinity Gamma Series Page 5-19
Message: V Fib
Parameter value: VF
Detection of a ventric-
ular fibrillation. The
ventricular rhythm is
chaotic.
Observe the patient and
treat if clinically indicated.
View each lead on the mon-
itor and check the ampli-
tude. To detect the QRS,
the amplitude should be at
least 0.5 mV on the mV
scale.
Reposition or change the
electrodes if the amplitude
is still low.
Use good skin preparation.
Message: V Tach
Parameter value: VT
n or more consecutive
PVCs have been
detected with a beat-
to-beat rate greater
than or equal to the
VT rate set in the
Alarm Tables. (n = VT
count parameter)
Observe the patient and
treat if clinically indicated.
Message: Bradycardia
Parameter value: BRDY
8 or more consecutive
QRS complexes have
been detected with a
beat-to-beat rate less
than 80% of the
patient’s normal rate.
Observe the patient and
treat if clinically indicated.
Message: RUN
Parameter value: RUN
3 to N-1* consecutive
PVCs with a beat-to-
beat rate >= the VT
rate
Observe the patient and
treat if clinically indicated.
Message: AIVR
Parameter value: AIVR
3 or more PVCs with a
rate < the VT rate
Observe the patient and
treat if clinically indicated.
Message: SVT
Parameter value: SVT
N* or more consecu-
tive normal beats with
a beat-to-beat rate >=
the SVT rate
Observe the patient and
treat if clinically indicated.
Message: CPT
Parameter value: CPT
A sequence of beats
with the pattern: nor-
mal, PVC, PVC, nor-
mal
Observe the patient and
treat if clinically indicated.
Message: BGM
Parameter value: BGM
A sequence of beats
with the pattern: nor-
mal, PVC, normal,
PVC, normal
Observe the patient and
treat if clinically indicated.
Heart Rate (HR), Arrhythmia (ARR) and ST Segment Analysis (ST)