Page 5-20 Infinity Gamma Series VF4
Message: TACH
Parameter value: TACH
N* or more consecu-
tive normal beats with
a beat-to-beat rate >=
the TACH rate
Observe the patient and
treat if clinically indicated.
Message: PAUS
Parameter value: PAUS
A sequence of two
normal beats with an
N-N* interval > the
Pause rate times the
average N-N interval
(±100ms)
Observe the patient and
treat if clinically indicated.
Message: ST <lead> Off
ST<lead1> and
ST<lead2> parameter
value: ***
The monitor has
detected a lead-off
condition for the cur-
rently processed lead.
The cause could be
one of the following:
Unplugged cable
Broken cable
Loose lead wire
Faulty lead wire
Dried out gel on the
electrode
Wrong cable type
selection.
The M
ULTIMED/
NEOMED pod may be
defective.
Check the patient cable and
lead wires carefully.
Check the MULTIMED pod.
Replace pod if defective.
Replace any cable or lead
wire that is suspect.
Reapply gel on the elec-
trode or change the elec-
trode (if disposable).
Select another ECG lead
for processing if the elec-
trode or lead cannot be
replaced.
Message: Cannot Analyze
ST<lead1> and
ST<lead2> parameter
value: <blanK>
ST cannot analyze
with a paced beat
ST segment analysis can-
not be performed on pace-
maker patients
Message: ST Too High
ST<lead1> and
ST<lead2> parameter
value: +++
Message: ST Too Low
ST<lead1> and
ST<lead2> parameter
value: ---
The detected ST
value is outside the
measurement range.
Check the patient and apply
treatment if necessary.
Check the electrode place-
ment and change their posi-
tion if necessary.
Heart Rate (HR), Arrhythmia (ARR) and ST Segment Analysis (ST)