Page 5-22 Infinity Gamma Series VF4
Respiration
Displayed Message Possible Cause Suggested Action
Message: Apnea
Parameter value: Apn
The monitor is
detecting an apnea
event that exceeds
the apnea time set in
the menu.
Incorrect placement
of the electrodes.
Improper threshold.
Missed shallow
breaths.
• Check the patient. Treat the
patient, if clinically indicated.
• If the absence of impedance
changes is due to the place-
ment of the electrodes, repo-
sition them until the
message clears.
• If apnea is due to an
improper threshold, set the
threshold until the message
clears.
• If breaths are missing, use
the manual respiration mode
to ensure the detection of
shallow breaths.
Message: MultiMed Dis-
connected
Parameter value: ***
The M
ULTIMED/
NEOMED pod is dis-
connected from the
monitor.
• Verify the pod connections.
Message: Coincidence
Parameter value: <value>
(Neonatal mode only)
The impedance
changes due to car-
diac activity are
being counted as
breaths.
• Check the patient.
• Check the electrode place-
ment.
Message: Rsp Fault
Parameter value: ***
Technical problem. • Turn the monitor off, then on.
• If the problem persists, take
the unit out of operation and
call DrägerService.
Message: Rsp High
Impedance
Parameter value: ***
Impedance out-of-
range.
Broken cables.
Dried out gel on the
electrodes.
Poor skin prepara-
tion.
• Check the cables for dam-
age.
• Check the electrodes.
• Move the electrodes, if nec-
essary.
Message: Rsp Testing
Parameter value:<blank>
Self-tests in
progress.
No displayed value.
• Wait until the end of the self-
tests.