Page 5-28 Infinity Gamma Series VF4
Message: etCO
2
Tubing
Blocked
Parameter value: ***
Sidestream tubing
obstructed, or filter is
clogged.
Clear the blockage in
the tubing.
Replace the pod.
Message: etCO
2
Tubing
Leak
Parameter value: ***
Sidestream tubing has a
leak.
Change the tubing.
Message: etCO
2
Hardware Failure
Parameter value: ***
etCO
2
communication or
hardware failure.
Check etCO
2
pod con-
nection.
Check all tubing
Disconnect the etCO
2
pod, then reconnect it.
If the message persists,
contact your Biomed.
Multigas
Displayed Message Possible Cause Suggested Action
Message: iCO
2
/etCO
2
Out of
Range
Parameter value: +++
The inspired/expired
concentrations fall
outside the monitor’s
display range.
Observe the patient;
treat, if necessary.
Check connections.
Disconnect and recon-
nect the Scio module.
Power-cycle the moni-
tor or undock/redock
the monitor from the
Docking Station.
Contact DrägerService.
Message: iO
2
/etO
2
Out of
Range
Parameter value: +++
Message: i/et <agent> Out of
Range
Parameter value: +++
End Tidal CO
2
(etCO
2
), Inspired CO
2
(iCO
2
), and RRc (continued)