Alarms and Messages
VF4 Infinity Gamma Series Page 5-29
Message: iO
2
> UL
Message: iO
2
< LL
Parameter value: <value>
The inspired/expired
concentrations fall
outside the current
upper or lower alarm
limits.
Observe the patient;
treat, if necessary.
Adjust alarm limits.
Message: etO
2
> UL
Message: etO
2
< LL
Parameter value: <value>
Message: i <agent> > UL
Message: i <agent> < LL
Parameter Value: <value>
Message: et <agent> > UL
Message: et <agent> < LL
Parameter value: <value>
Message: <agent> detected
Parameter value: <blank>
The user has specified
an agent (Agent Over-
ride), but Scio is
detecting a different
agent.
Turn Agent Override
off.
Message: Mixed Agents
Parameter value: ***
Scio is sensing more
than one agent in the
breathing circuit.
After a change of
agent, wait until all
traces of the first agent
have been cleared out
of the breathing circuit.
Message: <parameter>
reduced accuracy
Parameter value: <value>
Scio reports reduced
or unknown accuracy
even after completion
of warm-up period.
Observe the patient;
treat, if necessary.
Check connections.
Disconnect and recon-
nect the Scio module.
Power-cycle the moni-
tor or undock/redock
the monitor from the
Docking Station.
Contact DrägerService.
Message: Multigas
Unplugged
Parameter value: ***
No connection or
faulty connection
between monitor and
Scio module.
Check connections and
cables.
If message persists,
contact DrägerService.