Page 5-30 Infinity Gamma Series VF4
Message: Multigas H/W fail-
ure
Parameter value: ***
Loss of communica-
tion or hardware prob-
lem detected.
Check connections.
Disconnect and recon-
nect the Scio module.
Power-cycle the moni-
tor or undock/redock
the monitor from the
Docking Station.
Contact DrägerService.
Message: Multigas initializa-
tion
Parameter value: <blank>
Scio start-up mode. Informational message.
Message: Multigas warming
up
Parameter value: <values>
Scio warm-up period:
Reduced accuracy.
Displayed values might
not be accurate. Scio
reaches full accuracy
after a warm-up period
of about 5 minutes.
Message: Multigas zero in 1
minute
Parameter value: <values>
Scio will start the auto-
matic zeroing process
in one minute.
To delay zeroing for 5
minutes, click on
Autozero Delay in the
Multigas Setup menu
(see the chapter Multi-
gas).
Message: Multigas zero in
progress
Parameter value: <values>
Scio is undergoing
automatic zeroing pro-
cess. During zeroing
(< 25s), the monitor
does not update the
displayed gas values.
Wait until zeroing is
completed. During
zeroing, displayed gas
values might not be
accurate.
Message: Multigas zero OK
Parameter value: <values>
Scio zeroing was suc-
cessful.
Informational message.
Message: Multigas zero
failed
Parameter value: ***
Scio zeroing was not
successful.
Verify that the environ-
ment does not contain
contaminants.
Check for leaks or
occlusions.
Contact DrägerService.
Message: Multigas Tubing
Blocked
Parameter value: ***
Scio reports an occlu-
sion or pneumatic
problem.
Check all catheters for
occlusion.
If necessary, replace
catheters.
Multigas (continued)