Default Settings and Biomedical Support
VF4 Infinity Gamma Series Page C-3
Adult
Settings
Pediatric
Settings
Neonatal
Settings
After patient
discharge,
settings
return to...?
ECG
Alarm OnOnOn Saved Setup
Alarm Limits
(upper/lower)
120/45 150/50 180/80 Saved Setup
Alarm Recording Off Off Off Previous
Patient
Tone Source ECG ECG ECG Previous
Patient
Tone Volume Medium Medium Medium Previous
Patient
Pacer Detect Off Off N/A Saved Setup
QRS Marks Off Off Off Factory
ECG Processing ECG 1&2 ECG 1&2 ECG 1 Previous
Patient
ECG Leads 5-Lead 5-Lead 3-Lead Previous
Patient
Arrhythmia*
ASY, VF, VT,
RUN, AIVR, SVT,
CPT, BGM Alarm
BRDY, TACH,
PAUS, ARTF
Alarm
On
Off
On
Off
n/a
n/a
Saved Setup
ASY and VF
Alarm Recording
Other Events
Record
Off
Record
Off
n/a
n/a
Saved Setup
Arrhythmia
Monitoring
Basic Basic n/a Saved Setup
*Not applicable for neonatal monitoring.