Technical Data
VF4 Infinity Gamma Series Page D-5
Environmental requirements (Base Unit and AC Power Adapter)
Cooling Convection and cooling chimney (no fan)
Temperature Operating: 0 °C to +40 °C (without recorder)
Storage: –20 °C to 50 °C
Relative Humidity Operating: > 30% and < 95%, non-condensing
Storage: > 10% and < 95% non-condensing
Altitude Operating: -381 to +3048 m (-1250 to +10000 ft.)
525 to 795 mmHg (70 to 106 kPa)
Storage: -381 to +5486m (-1250 to +18000 ft.)
375 to 795 mmHg (50 to 106 kPa)
Transportation
(with shipping package)
Per National Safe Transit Association (NSTA)
Vibration 5-30 Hz 0.025DA, 30-500 Hz 1 G peak, 1 octave/min
sweep rate
14200 vibratory impacts with a total movement of 1"
Drop Per IEC68-2-31, Packaged drop- 30"
Shock 50 G peak, 11 ms duration, 3 positive and 3 negative
pulses in each axis.
Water Resistance Drip-Proof (IPX 1)
AC Power Adapter (59 53 539 E530U)
Power Requirements 100-120 VAC, 0.8A
200-240 VAC, 0.4A
Output Max. 34 W
Frequency 50-60 Hz
Chassis Leakage Current 300 µA @ 110 V and 60 Hz (UL 544)
500 µA @ 220 V and 50 Hz (IEC 601-1)
Mode of Operation Continuous
Protection Class Class I