Technical Data
VF4 Infinity Gamma Series Page D-11
R50 N Network Recorder
Dimensions 180 x 120 x 222 mm (7.1 x 4.72 x 8.74 in.)
Weight 1.64 kg (3.6 lb)
Connections AC Power Connector, Infinity Network, R50 Recorder,
Potential Equalization Connector.
Cooling Convection
Input voltage range 100-240 VRMS
Mains frequency 50/60 Hz
Power Consumption 1.0 A max
Protection Class Class I
Chassis Leakage Cur-
rent
<300 mA @ 120VAC, <500 mA @ 220VAC
Mode of Operation Continuous
Water Resistance Ordinary
Environmental Require-
ments
Temperature:
Operating: 15°C to 40°C (55°F to 104°F)
Storage: -20°C to 40°C (-4°F to 104°F)
Relative Humidity:
Operating: 30% to 95%, non-condensing
Storage: 10% to 95%, non-condensing
with packaging
Atmospheric Pressure:
Operating: 550 to 775 mmHg (73 to 103 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)