Technical Data
VF4 Infinity Gamma Series Page D-15
Measurement Range etCO
2
: 0-100 mmHg
iCO
2
: 0-10 mmHg
RRc: 5-145 bpm
Accuracy etCO
2
: ±2 mmHg for 0-40 mmHg
±5% reading for 41-70 mmHg
±8% reading for 71-100 mmHg
iCO
2
: ±2 mmHg
RRc: 1 bpm
Stable over 24 hours, over full range of readings at
atmospheric pressure.
Calibration Verify once a day. Calibrate when moving the sensor
from one module to another.
Calibration time: < 20 s
Compensation Gas/Anesthetic agent: User-selectable
Atm. pressure: Automatic or user-selectable
(540 - 800 mmHg)
Sampling Flow Rate 180 ±12 ml/min (Sidestream measuring mode)
Apnea Detection Yes
RRc Range (Pod) Mainstream: 0-149 breaths/min
Sidestream: 0-69 breaths/min
Accuracy: ±1 breath/min
Rise Time Mainstream: <100 ms
Sidestream: <200 ms
Delay Time Mainstream: <100 ms
Sidestream: <450 ms
Total System Response
Time
Rise time plus delay time
Resolution etCO
2
/iCO
2
: 1 mmHg, RRc: 1bpm
Average Response
Time
< 75ms from sensor
Breath Detection
Threshold
5 mmHg or greater
etCO
2
(continued)