Instructions for Use Oxylog 2000 plus SW 1.n 105
Technical Data
Settings
Ventilation modes VC-CMV, VC-AC, VC-SIMV, SpnCPAP,
Optional: Pressure support for VC-SIMV
and SpnCPAP.
Ventilation frequency RR 2 to 50 /min ±1 /min (VC-SIMV)
5 to 50 /min ±1 /min (VC-CMV, VC-AC)
12 to 50 /min ±1 /min for apnea ventilation.
Ventilation time ratio I:E (VC-CMV, VC-AC) 1:4 to 3:1
Inspiration time Ti (VC-SIMV, VC-SIMV / PS) 0.2 to 10 s
Inspiration tidal volume VT 0.1 to 2.0 L, BTPS
1)
Accuracy of Setting ±15% of set value or ±25 mL, whichever is greater.
O
2 concentration 100 vol.% (No-AirMix) or
approximately 40 vol.% (O
2 AirMix).
Positive end expiratory pressure PEEP 0 to 20 mbar ±2 mbar, no negative pressure.
Trigger sensitivity (flow trigger) 3 to 15 L/min
Pressure support Psupp 0 to 35 mbar (relative to PEEP) ±2 mbar
Rise time for pressure
support
slow (1 s), standard (0.4 s), fast (0 s).
1) BTPS
Body Temperature, Pressure, Saturated.
Measured values referred to the conditions of the patient's lungs,
body temperature 37
o
C, airway pressure, water-vapor-saturated gas.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plu and is the answer not in the manual?

Dräger Oxylog 2000 plu Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plu
CategoryMedical Equipment
LanguageEnglish

Related product manuals