Technical Data
116
Instructions for Use Oxylog 2000 plus SW 1.n
Recommended separation distances
Recommendated separation distances between
portable and mobile RF-Telecommunication
devices and the Oxylog 2000 plus.
The table below gives examples of the most com-
mon devices. If a device is not listed, do as follows:
1 Check the product manual of the device for the
energy (W) and the frequency (GHz) values that
the device transmits.
2 In the table below, search for the energy (W)
value of the device in the PEIRP (W) column.
3 On the correct table row, search for the distance
that corresponds to the frequency of the device.
If the frequency is between 150 kHz to 2.5 GHz,
use the 3 V/m distance column.
Otherwise, use the 1 V/m distance column.
Recommended separation distances between portable and mobile RF-Telecommunication
devices and the Oxylog 2000 plus
maximum
PEIRP (W)
3 V/m
distance (m)
1 V/m
distance (m)
Examples
0.001 0.06 0.17
0.003 0.10 0.30
0.010 0.18 0.55 e.g. Garage door openers
0.030 0.32 0.95 e.g. WLAN 5250 / 5775 (Europe)*
0.100 0.58 1.73 e.g. WLAN 2440 (Europe), Bluetooth*
0.200 0.82 2.46 e.g. WLAN 5250 (not in Europe)*
0.250 0.91 2.75 e.g. DECT devices*
1.000 1.83 5.48 e.g. GSM 1800- / GSM 1900- / UMTS- mobiles,
WLAN 5600 (not in Europe)*
2.000 2.60 7.78 e.g. GSM 900 mobiles*
3.000 3.16 9.49
Information regarding separation distances (IEC 60601-1-2, tables 205 and 206).
* Telecommunication devices. For the correct type, check the product manual of the device.