Technical Data
106
Instructions for Use Oxylog 2000 plus SW 1.n
Performance data
Control principle Time-cycled, volume-constant, pressure supported.
Maximum inspiratory flow 100 L/min
1)
Device compliance
with 1.5 m ventilation hose< 1 mL/mbar
with 3 m ventilation hose <
2 mL/mbar
Inspiration resistance < 6 mbar at 60 L/min
< 4 mbar at 30 L/min
<
2 mbar at 5 L/min
Expiration resistance <
6 mbar at 60 L/min
< 4 mbar at 30 L/min
<
2 mbar at 5 L/min
Dead space including flow sensor approximately 35 mL (reusable hose system).
approximately 33 mL (disposable hose system).
Supplementary functions
Demand valve Opens the breathing system upon failure of the gas
supply, permits spontaneous breathing with ambient
air.
Relief valve Opens the breathing system at
approximately 80 mbar.
Patient connection 22 mm ISO conical connector.
1) At service pressures >350 kPa.
The maximum inspiratory flow is reduced to 80 L/min at service pressures <350 kPa and to 39 L/min at service
pressures <280 kPa.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plu and is the answer not in the manual?

Dräger Oxylog 2000 plu Specifications

General IconGeneral
CategoryMedical Equipment
TypeTransport Ventilator
Tidal Volume50 to 2000 mL
PEEP0 to 20 cmH2O
DisplayMonochrome LCD
Inspiratory Time0.2 to 5 s
FiO221 to 100%
Power Supply100 to 240 V AC, 50/60 Hz
Ventilation ModesIPPV, SIMV, CPAP

Related product manuals