Checking readiness for operation
Whenever the ventilator has been serviced or
the ventilation hoses changed.
At the latest every six months.
The following functions are checked with the menu-
based test:
Gas supply present.
Hose system / breathing valve connected and
OK.
Alarm functions OK.
Ventilation functions OK.
Monitor functions OK.
Oxylog 2000 plus interrupts the test if a fault is
detected.
The relevant fault is indicated on the screen.
WARNING
The patient may be endangered if the above
device check is not completed.